Vibrational circular dichroism
Vibrational circular dichroism is a spectroscopic technique which detects differences in attenuation of left and right circularly polarized light passing through a sample. It is the extension of circular dichroism spectroscopy into the infrared and near infrared ranges.
Because VCD is sensitive to the mutual orientation of distinct groups in a molecule, it provides three-dimensional structural information. Thus, it is a powerful technique as VCD spectra of enantiomers can be simulated using ab initio calculations, thereby allowing the identification of absolute configurations of small molecules in solution from VCD spectra. Among such quantum computations of VCD spectra resulting from the chiral properties of small organic molecules are those based on density functional theory and gauge-including atomic orbitals. As a simple example of the experimental results that were obtained by VCD are the spectral data obtained within the carbon-hydrogen stretching region of 21 amino acids in heavy water solutions. Measurements of vibrational optical activity have thus numerous applications, not only for small molecules, but also for large and complex biopolymers such as muscle proteins and DNA.
Vibrational modes
Theory
While the fundamental quantity associated with the infrared absorption is the dipole strength, the differential absorption is also proportional to the rotational strength, a quantity which depends on both the electric and magnetic dipole transition moments. Sensitivity of the handedness of a molecule toward circularly polarized light results from the form of the rotational strength. A rigorous theoretical development of VCD was developed concurrently by the late Professor P.J. Stephens, FRS, at the University of Southern California, and the group of Professor A.D. Buckingham, FRS, at Cambridge University in the UK, and first implemented analytically in the Cambridge Analytical Derivative Package by R.D. Amos. Previous developments by D.P. Craig and T. Thirmachandiman at the Australian National University and Larry A. Nafie and Teresa B. Freedman at Syracuse University though theoretically correct, were not able to be straightforwardly implemented, which prevented their use. Only with the development of the Stephens formalism as implemented in CADPAC did a fast efficient and theoretically rigorous theoretical calculation of the VCD spectra of chiral molecules become feasible. This also stimulated the commercialization of VCD instruments by Biotools, Bruker, Jasco and Thermo-Nicolet.Peptides and proteins
Extensive VCD studies have been reported for both polypeptides and several proteins in solution; several recent reviews were also compiled. An extensive but not comprehensive VCD publications list is also provided in the "References" section. The published reports over the last 22 years have established VCD as a powerful technique with improved results over those previously obtained by visible/UV circular dichroism or optical rotatory dispersion for proteins and nucleic acids.The effects due to solvent on stabilizing the structures of amino acids and peptides and the corresponding effects seen in the vibrational circular dichroism and Raman optical activity spectra have been recently documented by a combined theoretical and experimental work on L-alanine and N-acetyl L-alanine N'-methylamide. Similar effects have also been seen in the nuclear magnetic resonance spectra by the Weise and Weisshaar NMR groups at the University of Wisconsin–Madison.
Nucleic acids
VCD spectra of nucleotides, synthetic polynucleotides and several nucleic acids, including DNA, have been reported and assigned in terms of the type and number of helices present in A-, B-, and Z-DNA.Instrumentation
VCD can be regarded as a relatively recent technique. Although Vibrational Optical Activity and in particular Vibrational Circular Dichroism, has been known for a long time, the first VCD instrument was developed in 1973 and commercial instruments were available only since 1997.For biopolymers such as proteins and nucleic acids, the difference in absorbance between the levo- and dextro- configurations is five orders of magnitude smaller than the
corresponding absorbance. Therefore, VCD of biopolymers requires the use of very sensitive, specially built instrumentation as well as time-averaging over relatively long intervals of time even with such sensitive VCD spectrometers.
Most CD instruments produce left- and right- circularly polarized light which is then either sine-wave or square-wave modulated, with subsequent phase-sensitive detection and lock-in amplification of the detected signal. In the case of FT-VCD,
a photo-elastic modulator is employed in conjunction with an FTIR interferometer set-up. An example is that of a Bomem model MB-100 FTIR interferometer equipped with additional polarizing optics/ accessories needed for recording VCD spectra.
A parallel beam emerges through a side port of the interferometer which passes first through a wire grid linear polarizer and then through an octagonal-shaped ZnSe crystal PEM which modulates the polarized beam at a fixed, lower frequency such as 37.5 kHz. A mechanically stressed crystal such as ZnSe exhibits birefringence when stressed by an adjacent piezoelectric transducer. The linear polarizer is positioned close to, and at 45 degrees, with respect to the ZnSe crystal axis. The polarized radiation focused onto the detector is doubly modulated, both by the PEM and by the interferometer setup. A very low noise detector, such as MCT, is also selected for the VCD signal phase-sensitive detection. The first dedicated VCD spectrometer brought to market was the ChiralIR from Bomem/BioTools, Inc. in 1997. Today, Thermo-Electron, Bruker, Jasco and BioTools offer either VCD accessories or stand-alone instrumentation. To prevent detector saturation an appropriate, long wave pass filter is placed before the very low noise MCT detector, which allows only radiation below 1750 cm−1 to reach the MCT detector; the latter however measures radiation only down to 750 cm−1. FT-VCD spectra accumulation of the selected sample solution is then carried out, digitized and stored by an in-line computer. Published reviews that compare various VCD methods are also available.
Modeling VCD Spectra
In 1994, researchers at the University of Southern California, U.S. Army Research Laboratory, and Lorentzian Inc., reported an accuracy ranking of quantum mechanical analytical techniques to theoretically determine vibrational frequencies, dipole strengths, and rotational strengths of an organic molecule. This ranking claimed density functional theory at the B3LYP/6-31G* level of theory was the most accurate and effective computation used to model and vibrational circular dichroism spectra. Electronic structure computations, by solving either the Schrödinger equation or the Kohn–Sham equation, can be used to obtain information about ground state energy, bond vibrational frequency, and electron density, and other characteristics.Theoretical calculations of vibrational energy often involve the Schrödinger's equation with the Hamiltonian operator. The computers that process these massive calculations can incorporate the molecule's kinetic energy as well as the vast number of repulsions and Coulombic attractions between subatomic particles. The calculations are said to be very costly, as they are difficult and take a long time to accomplish. This is partially because integrating the electron-electron interactions into the equation involves determining electron exchange interactions. Methods like DFT and the Hartree–Fock method look at a group of atomic orbitals referred to as a basis set to estimate the molecular wave function. The wave function can be used to calculate chiroptical properties of interest such as frequency, wavelength, energy etc. Hartree–Fock operates with a feedback loop called a self-consistent field that continuously refines the wave function estimates until the value falls within a satisfactory change in energy threshold that the calculation converges to an approximate wavefunction solution.
The study conducted by USC, USARL, and Lorentzian Inc. analyzed the infrared and VCD spectra of the chiral molecule 4-methyl-2-oxetanone. Lorentzian bands were fit to FTIR and VCD spectra to obtain their peak intensity, line width, and frequency, which may be used to infer properties such as dipole strengths and rotational strengths. These experimental values were then compared to theoretical results. The scientists reported that DFT computations evaluated with the B3LYP functional best modelled FTIR and VCD spectra. To achieve a better cost to benefit ratio the researchers recommended pairing this method with the 6-31G* basis set. The second best method reported was the second order Møller–Plesset perturbation theory. The third and fourth best calculation methods were DFT with the BLYP and LSDA functionals respectively. The researchers stated that Ab initio Hartree–Fock Self-Consistent Field computations modelled FTIR and VCD spectra with the lowest accuracy compared to other methodologies investigated.
The significance in claimed improvement in accuracy of DFT computations over ab initio techniques was that DFT computations were reported to quicken computational speed. By evaluating an effective potential using electron density, which can by specified across three degrees of freedom, DFT sidesteps the evaluation of coulombic potentials between every single electron, which is specified over 3N degrees of freedom. The B3LYP basis set is a hybrid between direct Hartree–Fock exchange terms as well as local and gradient corrections for exchange and correlation interactions. Therefore, the B3LYP functional is claimed to efficiently model FTIR and VCD of some molecules via DFT at a fraction of the cost.