Thin-layer chromatography
Thin-layer chromatography is a chromatography technique used to separate non-volatile mixtures. Thin-layer chromatography is performed on a sheet of glass, plastic, or aluminium foil, which is coated with a thin layer of adsorbent material, usually silica gel, aluminium oxide, or cellulose. This layer of adsorbent is known as the stationary phase.
After the sample has been applied on the plate, a solvent or solvent mixture is drawn up the plate via capillary action. Because different analytes ascend the TLC plate at different rates, separation is achieved. The mobile phase has different properties from the stationary phase. For example, with silica gel, a very polar substance, non-polar mobile phases such as heptane are used. The mobile phase may be a mixture, allowing chemists to fine-tune the bulk properties of the mobile phase.
After the experiment, the spots are visualized. Often this can be done simply by projecting ultraviolet light onto the sheet; the sheets are treated with a phosphor, and dark spots appear on the sheet where compounds absorb the light impinging on a certain area. Chemical processes can also be used to visualize spots; anisaldehyde, for example, forms colored adducts with many compounds, and sulfuric acid will char most organic compounds, leaving a dark spot on the sheet.
To quantify the results, the distance traveled by the substance being considered is divided by the total distance traveled by the mobile phase. This ratio is called the retardation factor. In general, a substance whose structure resembles the stationary phase will have low Rf, while one that has a similar structure to the mobile phase will have high retardation factor. Retardation factors are characteristic, but will change depending on the exact condition of the mobile and stationary phase. For this reason, chemists usually apply a sample of a known compound to the sheet before running the experiment.
Thin-layer chromatography can be used to monitor the progress of a reaction, identify compounds present in a given mixture, and determine the purity of a substance. Specific examples of these applications include: analyzing ceramides and fatty acids, detection of pesticides or insecticides in food and water, analyzing the dye composition of fibers in forensics, assaying the radiochemical purity of radiopharmaceuticals, or identification of medicinal plants and their constituents
A number of enhancements can be made to the original method to automate the different steps, to increase the resolution achieved with TLC and to allow more accurate quantitative analysis. This method is referred to as HPTLC, or "high-performance TLC". HPTLC typically uses thinner layers of stationary phase and smaller sample volumes, thus reducing the loss of resolution due to diffusion.
Plate preparation
TLC plates are usually commercially available, with standard particle size ranges to improve reproducibility. They are prepared by mixing the adsorbent, such as silica gel, with a small amount of inert binder like calcium sulfate and water. This mixture is spread as a thick slurry on an unreactive carrier sheet, usually glass, thick aluminum foil, or plastic. The resultant plate is dried and activated by heating in an oven for thirty minutes at 110 °C. The thickness of the absorbent layer is typically around 0.1–0.25 mm for analytical purposes and around 0.5–2.0 mm for preparative TLC.Technique
The process is similar to paper chromatography with the advantage of faster runs, better separations, and the choice between different stationary phases. Because of its simplicity and speed, TLC is often used for monitoring chemical reactions and for the qualitative analysis of reaction products. Plates can be labeled before or after the chromatography process using a pencil or other implement that will not interfere or react with the process.To run a thin layer chromatography plate, the following procedure is carried out:
- Using a capillary tube, a small spot of solution containing the sample is applied to a plate, about 1.5 centimeters from the bottom edge. The solvent is allowed to completely evaporate off to prevent it from interfering with sample's interactions with the mobile phase in the next step. If a non-volatile solvent was used to apply the sample, the plate needs to be dried in a vacuum chamber. This step is often repeated to ensure there is enough analyte at the starting spot on the plate to obtain a visible result. Different samples can be placed in a row of spots the same distance from the bottom edge, each of which will move in its own adjacent lane from its own starting point.
- A small amount of an appropriate solvent is poured into a glass beaker or any other suitable transparent container to a depth of less than 1 centimeter. A strip of filter paper is put into the chamber so that its bottom touches the solvent and the paper lies on the chamber wall and reaches almost to the top of the container. The container is closed with a cover glass or any other lid and is left for a few minutes to let the solvent vapors ascend the filter paper and saturate the air in the chamber.
- The TLC plate is then placed in the chamber so that the spot of the sample do not touch the surface of the eluent in the chamber, and the lid is closed. The solvent moves up the plate by capillary action, meets the sample mixture and carries it up the plate. The plate should be removed from the chamber before the solvent front reaches the top of the stationary phase and dried.
- Without delay, the solvent front, the furthest extent of solvent up the plate, is marked.
- The plate is visualized. As some plates are pre-coated with a phosphor such as zinc sulfide, allowing many compounds to be visualized by using ultraviolet light; dark spots appear where the compounds block the UV light from striking the plate. Alternatively, plates can be sprayed or immersed in chemicals after elution. Various visualising agents react with the spots to produce visible results.
Separation process and principle
displaying thin-layer chromatography. The horizontal blue strip is from a reaction between the iron bandsaw supports and the acidic timber.
Separation of compounds is based on the competition of the solute and the mobile phase for binding sites on the stationary phase. For instance, if normal-phase silica gel is used as the stationary phase, it can be considered polar. Given two compounds that differ in polarity, the more polar compound has a stronger interaction with the silica and is, therefore, better able to displace the mobile phase from the available binding sites. As a consequence, the less polar compound moves higher up the plate. If the mobile phase is changed to a more polar solvent or mixture of solvents, it becomes better at binding to the polar plate and therefore displacing solutes from it, so all compounds on the TLC plate will move higher up the plate. It is commonly said that "strong" solvents push the analyzed compounds up the plate, whereas "weak" eluents barely move them. The order of strength/weakness depends on the coating of the TLC plate. For silica gel-coated TLC plates, the eluent strength increases in the following order: perfluoroalkane, hexane, pentane, carbon tetrachloride, benzene/toluene, dichloromethane, diethyl ether, ethyl acetate, acetonitrile, acetone, 2-propanol/n-butanol, water, methanol, triethylamine, acetic acid, formic acid. For C18-coated plates the order is reverse. In other words, when the stationary phase is polar and the mobile phase is nonpolar, the method is normal-phase as opposed to reverse-phase. This means that if a mixture of ethyl acetate and hexane as the mobile phase is used, adding more ethyl acetate results in higher R''f values for all compounds on the TLC plate. Changing the polarity of the mobile phase will normally not result in reversed order of running of the compounds on the TLC plate. An eluotropic series can be used as a guide in selecting a mobile phase. If a reversed order of running of the compounds is desired, an apolar stationary phase should be used, such as C18-functionalized silica.
Analysis
As the chemicals being separated may be colorless, several methods exist to visualize the spots:- Fluorescent analytes, like quinine, may be detected under blacklight
- Often a small amount of a fluorescent compound, usually manganese-activated zinc silicate, is added to the adsorbent that allows the visualization of spots under UV-C light. The adsorbent layer will thus fluoresce light-green by itself, but spots of analyte quench this fluorescence.
- Iodine vapors are a general unspecific color reagent
- Specific color reagents into which the TLC plate is dipped or which are sprayed onto the plate exist.
- *Potassium permanganate – oxidation
- * Bromine
- * Acidic vanillin
- *Phosphomolybdic acid
- In the case of lipids, the chromatogram may be transferred to a polyvinylidene fluoride membrane and then subjected to further analysis, for example mass spectrometry, a technique known as far-eastern blot.
Applications
Characterization
In organic chemistry, reactions are qualitatively monitored with TLC. Spots sampled with a capillary tube are placed on the plate: a spot of starting material, a spot from the reaction mixture, and a cross-spot with both. A small TLC plate takes a couple of minutes to run. The analysis is qualitative, and it will show if the starting material has disappeared, i.e. the reaction is complete, if any product has appeared, and how many products are generated. Unfortunately, TLCs from low-temperature reactions may give misleading results, because the sample is warmed to room temperature in the capillary, which can alter the reaction—the warmed sample analyzed by TLC is not the same as what is in the low-temperature flask. One such reaction is the DIBALH reduction of ester to aldehyde.In one study TLC has been applied in the screening of organic reactions, for example in the fine-tuning of BINAP synthesis from 2-naphthol. In this method, the alcohol and catalyst solution are placed separately on the baseline, then reacted, and then instantly analyzed.
A special application of TLC is in the characterization of radiolabeled compounds, where it is used to determine radiochemical purity. The TLC sheet is visualized using a sheet of photographic film or an instrument capable of measuring radioactivity. It may be visualized using other means as well. This method is much more sensitive than the others and can be used to detect an extremely small amount of a compound, provided that it carries a radioactive atom.
Isolation
Since different compounds will travel a different distance in the stationary phase, chromatography can be used to isolate components of a mixture for further analysis. The separated compounds each occupying a specific area on the plate, they can be scraped off and dissolved into an appropriate solvent. As an example, in the chromatography of an extract of green plant material shown in 7 stages of development, Carotene elutes quickly and is only visible until step 2. Chlorophyll A and B are halfway in the final step and lutein the first compound staining yellow. Once the chromatography is over, the carotene can be removed from the plate, extracted into a solvent and placed into a spectrophotometer to determine its spectrum. The quantities extracted are small and a technique such as column chromatography is preferred to separate larger amounts. However, big preparative TLC plates with thick silica gel coatings can be used to separate more than 100 mg of material.Examining reactions and compound stability
TLC is also used for the identification of the completion of any chemical reaction. To determine this it is observed that at the beginning of a reaction the entire spot is occupied by the starting chemicals or materials on the plate. As the reaction starts taking place the spot formed by the initial chemicals starts reducing and eventually replaces the whole spot of starting chemicals with a new product present on the plate. The formation of an entirely new spot determines the completion of a reaction.Furthermore, two-dimensional TLC is frequently used as a method to check if a compound is stable in the stationary phase. For this purpose, the tested compound mixture is eluted twice in a square-shaped TLC plate, first in one direction and then rotated 90º. If the target compound appears on the diagonal of the square, it is stable in silica gel and safe to purify. If it appears below the diagonal, it is decomposing on silica gel. If this is the case, purification can be attempted using neutralized silica gel, or an alternative stationary phase such as neutral alumina.