Fluid-attenuated inversion recovery


Fluid-attenuated inversion recovery is an MRI sequence with an inversion recovery set to null fluids. For example, it can be used in brain imaging to suppress cerebrospinal fluid effects on the image, so as to bring out the periventricular hyperintense lesions, such as multiple sclerosis plaques. It was invented by Dr. Graeme Bydder. FLAIR can be used with both three-dimensional imaging or two dimensional imaging.

Technique

By carefully choosing the inversion time, the signal from any particular tissue can be nulled. The appropriate TI depends on the tissue via the formula:
in other words, one should typically use a TI of around 70% of the T1 value. In the case of CSF suppression, one aims for T1-weighted images, which prioritize the signal of fat over that of water. Therefore, if the long TI is adjusted to a zero crossing point for water, the signal of the CSF is theoretically being "erased," from the derived image.

Clinical applications

The FLAIR sequence analysis has been especially useful in the evaluation and study of CNS disorders, involving:
* Post-contrast FLAIR images have been added to diagnosis protocol for accurate medical assessment.